1---10.4” TFT LCD screen displays the Ventilation parameters, Alarming information and Waveform.
2---Recycle breathing pipe, ensure easy operating and keep tidy.
3---Multiple working modes such as volume control and pressure limit, adapt to wide range patient.
4---Vaporizer with temperature, flow compensation and self-lock function, keep safety anytime.
5---Multiple parameters monitoring interface, make every parameter clear, let users know the patient conditions in all aspects;
6---Real time pressure-time, flow-time loop graphics and high precision O2 concentration detection function included.
7---Venturi effect technology ensure the machine working without air gas source.
Specifications | |
Ventilation modes: | IPPV; A/C; SIMV; SPONT/CPAP; MANU |
Alarm and protection | |
The AC power failure alarm | Power failure or no connection |
Internal battery backup low voltage alarm | <11.3±0.3V |
No tidal volume | No tidal volume within 6s |
High oxygen concentration alarm | 19%-100% |
Low oxygen concentration alarm | 18%-99% |
High Airway pressure alarm | 20cmH2O-100cmH2O |
Low Airway pressure alarm | 0cmH2O-20cmH2O |
High Minute Volume alarm | 1L/min-99L/min |
Low Minute Volume alarm | 0-20L/min |
Continuous Pressure alarm | (PEEP+1.5kPa) over 16s |
Suffocation warning | 5s-60s no spontaneous ventilation |
Working conditions | |
Gas source | O2 |
Pressure | 280kPa-600kPa |
Voltage | -220V±22V |
Power frequency | 50Hz±1Hz |
Input power | 40VA |