Features
l 12.1” TFT LCD screen displays the Ventilation parameters, Alarming information and Waveform.
l Recycle breathing pipe, ensure easy operating and keep tidy.
l Multiple working modes such as volume control and pressure limit, adapt to wide range patient.
l Vaporizer with temperature, flow compensation and self-lock function, keep safety anytime.
l Multiple parameters monitoring interface, make every parameter clear, let users know the patient conditions in all aspects;
l Real time pressure-time, flow-time loop graphics and high precision O2 concentration detection function included.
l Stable and low noise air compressor, create quiet work environment for doctors.
Ventilation modes: | IPPV, A/C, PCV, SIMV, SPONT/CPAP,PSV,SIGH,MANU |
Ventilator parameter ranges | |
Tidal volume(Vt) | 0 ~2000mL |
Frequency (Freq) | 1bpm ~100 bpm |
Oxygen concentration | 21%-100% |
I/E | 4:1~1:8 |
PEEP | 0cmH2O ~40 cmH2O |
Pressure Limit | 20 cmH2O ~100 cmH2O |
Monitored Parameters | |
Frequency (Freq) | 0 /min ~100 /min |
Tidal volume(Vt) | 0 mL ~2500 mL |
Oxygen concentration | 15 % ~100 % |
Oscillographs display: | |
P-T(pressure – time) | |
F-T(flow – time) | |
P-V loop (pressure – volume loop) | |
Size | |
1st Wooden case packing size(main engine):L560*W 560*H 605mm , GW:35KG ; NW:17KG | |
2nd Wooden case packing size(air compressor):L670*W700*H 1160 mm , GW:79KG ; NW:46.2KG | |
Alarm and protection | |
The AC power failure alarm | Power failure or no connection |
Internal battery backup low voltage alarm | ≤11.3±0.3V |
No tidal volume | No tidal volume within 6s |
High Minute Volume alarm | 5L/min-99L/min |
LowMinute Volume alarm | 1L/min-30L/min |
High Airway pressure alarm | 20cmH2O-100cmH2O |
Low Airway pressure alarm | 0cmH2O-20cmH2O |
High oxygen concentration alarm | 19%-100% |
Low oxygen concentration alarm | 18%-99% |
Continuous pressure alarm | (PEEP+1.5cmH2O) over 16s |
Suffocation warning | 5-60s |
Fan error | Show on screen |
Oxygen deficit | Show on screen |
The maximum limited pressure | <12.5 kPa |
Working conditions | |
Gas source | O2,AIR |
Pressure | 280kPa-600kPa |
Voltage | -220V±22V |
Power frequency | 50Hz±1Hz |
Input power | 900VA(with air compressor) |
250VA(without air compressor) |